Model note · Infusion Pump

CareFusion BD Alaris System

The BD Alaris System is a modular pump: one PC Unit (8015) carries the screen, power and network, and up to four modules clip onto its sides (large volume Pump Module 8100, Syringe 8110, PCA 8120, EtCO2), plus an Auto-ID barcode module as a fifth. Guardrails drug libraries set dose limits. After many recalls, BD gained a new 510(k), K211218, on 21 July 2023 for updated hardware and v12 software with stronger cybersecurity. Most failures in the recall record are mechanical: the pump module bezel, keypad fluid ingress and the inter unit connectors.

The model

Maker, model and type.

Maker
CareFusion 303, Inc. (BD)
Model
BD Alaris System with Guardrails Suite MX v12 (PC Unit 8015 with Pump Module 8100)
Device type
Infusion Pump
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA classificationInfusion pump, 21 CFR 880.5725, Class II, product codes FRN (Pump), FRN syringe module, MEA (PCA) and CCK (EtCO2 analyser)Source
ClearancesK091308 (2009, PC Unit 8000 and 8015), K133532 (2014, Guardrails Suite MX), K211218 (21 July 2023, v12.1.2), K243855 (25 April 2025)Source
Module rulesUp to four infusion or monitoring modules per PC Unit, plus the Auto-ID Module as a fifth; only one PCA and one EtCO2 module; spread modules evenly on both sides for balance on the poleSource
Rate rangePump and PCA Modules 0.1 to 999 mL/h; Syringe Module 0.01 to 999 mL/hSource
Pump Module accuracy (v12)System flow accuracy -19% to +5.5% at 1 to 999 mL/h and -8% to +5.5% at 0.1 to 1 mL/h, tested to AAMI TIR101:2021 conditions (the older label said plus or minus 5%)Source
Maximum infusion pressurePump Module 525 mmHg; Syringe Module about 800 mmHg without the pressure sensing disc and 1060 mmHg with itSource
Power100 to 240 V AC, 50/60 Hz, 150 VA maximum; rechargeable 12 V, 4000 mAh NiMH battery in the PC UnitSource
Protection and environmentIPX2 (v12; the earlier system was IPX1); operating pressure 525 to 795 mmHg, so not for hyperbaric chambers; maximum supported configuration 13 kgSource
ContraindicationNot for the enteral route (v12)Source

Recalls

Known FDA recalls.

  • Z-1217-20262025-11-06

    Pump Modules 8100 flashed with a daylight saving date and time that can break hospital network interoperability.

    FDA record
  • Z-0430-20262025-10-17

    Dropped or jarred Pump Modules can have a damaged bezel, causing over or under infusion, free flow or failure to calibrate.

    FDA record
  • Z-2179-20252025-07-08

    Pump Module 8100 may perform outside its flow, bolus and occlusion time to alarm ranges with compatible sets.

    FDA record
  • Z-1231-20252025-01-23

    User manual wrongly instructed leaving the safety clamp open when reloading a set.

    FDA record
  • Z-3138-20242024-08-15

    PC Units 8015 with bent female IUI connectors may not work, delaying infusion or monitoring.

    FDA record
  • Z-1192-20232023-01-24

    PC Units with certain software, network cards and IP addresses lose wireless connection and cannot receive new datasets.

    FDA record
  • Z-2909-20202020-08-04

    Keypad keys stuck or unresponsive after fluid ingress, delaying or interrupting infusion.

    FDA record

Safety

Safety warnings from the record.

  • An earlier user manual wrongly said to leave the safety clamp open when reloading a set; with the roller clamp also open this can cause accidental over infusion. Source
  • In 2025 BD warned that Pump Modules used with compatible sets may perform outside their stated flow, bolus and occlusion alarm ranges. Source
  • The PC Unit and modules cannot be started remotely; only infusion parameters can be sent from the EHR and a clinician confirms at the pump. Source

Sources

Where this note comes from.

Not covered: The Alaris technical service manual is not public, so calibration steps, part numbers and the full system error list are not included; the error codes above are only those named in FDA recall records. There are more than 30 recall events for the 8015 and 8100 since 2013; seven recent ones are listed. The K243855 summary (2025) was downloaded but its changes are not yet folded in.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.