Andromeda Ellipse urodynamic system

The Andromeda Ellipse is a compact, mini trolley urodynamic system made in Germany and first sold in 1996. It records pressure, uroflow and EMG, converts the signals to digital form, shows them on a built in LCD screen, prints on a thermal printer and stores studies in memory for later transfer to a PC over a network. Pressures are taken with microtip (catheter tip) transducers rather than water filled lines. It supports flowmetry, cystometry, pressure flow studies, urethral pressure profiles and EMG. The original manufacturer website is no longer online, so support now depends on distributors.

The model

Maker, model and type.

Maker
Andromeda medizinische Systeme GmbH (Taufkirchen, Germany)
Model
Ellipse urodynamic system
Device type
Urodynamics System

Key specifications

What the public record says.

ItemValueSource
Tests supportedFlowmetry, cystometry, pressure flow, urethral pressure profile and EMGSource
Pressure sensingMicrotip catheters (MTCs) for pressure; surface electrodes for EMGSource
Maximum pressure channels8 (NHS CEP 2009 table)Source
Uroflowmeter typeWeight sensorSource
Monitor size15 cmSource
Display, printing and storageLCD screen, thermal printer and internal memory; data can be downloaded to a PC over a networkSource
Options listed in 2009Profilometer, EMG amplifier, video capture and user calibrationSource
Report and export formatsPDF and Word reports, text and ICS export, all via laptopSource
Configuration and warranty (UK, 2009)Mini trolley; 2 year warranty; supplied in the UK by MediplusSource
Pump and power (reseller listing for a refurbished unit, not manufacturer data)H2O pump 1 to 200 ml/min; 230 V, 50/60 Hz, 160 W; 22 kgSource
FDA clearanceK964522 for Ellipse 4, 21 CFR 876.1620, Class II, cleared 21 March 1997Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Reading from a catheter tip sensor depends on where the tip lies; the top of the bladder can read about 10 cmH2O lower than the bottom, and this cannot easily be corrected. Source
  • Transducers calibrated at room temperature must stay accurate at body temperature, so check readings after the catheter has warmed in the patient. Source
  • Denser fluids such as contrast medium make a weight based flowmeter over read flow and volume unless the density setting is changed. Source
  • Electrical safety tests and repairs should only be done by competent technicians. Source

Sources

Where this note comes from.

Not covered: Sources are thin. The former manufacturer website (andromeda-ms.de) no longer hosts Andromeda content, and no operator or service manual was found, so there are no error codes, part numbers or model specific calibration routines; first checks and failing parts rely on general ICS and NHS urodynamic guidance. The FDA 510(k) covers the earlier Ellipse 4 and is a scanned 1997 summary without numeric specs. Channel count, display and options come from a 2009 NHS evaluation. Pump, power and weight come only from a used equipment reseller and need confirming. No FDA recalls or MAUDE reports were found for Andromeda.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.