ACIST Kodama HD-IVUS catheter

ACIST's HDi system and Kodama catheter make up a rotational high definition IVUS platform. A spinning transducer at the catheter tip can run at 40 or 60 MHz to give cross sectional images of coronary and peripheral vessels. The system has a console, a handheld patient interface module (PIM) that spins the imaging core, and an optional linear translation system for automated pullback. It images by mechanical rotation rather than with a solid state array.

The model

Maker, model and type.

Maker
ACIST Medical Systems, Inc.
Model
ACIST Kodama HD-IVUS catheter with ACIST HDi System
Device type
IVUS
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Imaging frequencies40 MHz (standard definition) and/or 60 MHz (high definition)Source
Field of view6, 8 and 10 mm, extended in 2020 to 12, 16 and 20 mm; the system picks the highest definition available for the chosen fieldSource
Pullback and frame rate examples1.0 mm/s at 30 frames/s and 10.0 mm/s at 60 frames/s were tested at 20 mm field of viewSource
System componentsConsole with touchscreen, handheld PIM, optional linear translation system (LTS), Kodama catheterSource
Signal pathCoaxial cable embedded in the drive cable carries signals from the transducer back to the consoleSource
FDA classification and clearancesIYO (21 CFR 892.1560) and OBJ (21 CFR 870.1200), Class II; K173063 (2017), K191175 (2019), K193183 (decided 20 March 2020)Source

Recalls

Known FDA recalls.

  • Z-1161-20212021-01-22

    Damaged o-ring fragments over 200 microns could be flushed out of Kodama catheters; affected lots removed.

    FDA record

Safety

Safety warnings from the record.

  • Particles from a damaged internal o-ring could enter the bloodstream during flushing. Source
  • The 2020 software change introduced one new use related risk from the field of view and frequency selection, though the overall risk profile was judged unchanged. Source

Sources

Where this note comes from.

Not covered: No ACIST product page, catheter IFU or HDi operator manual was opened, so catheter size, length, guide catheter compatibility and system error messages are not included and manufacturerUrl points to the FDA summary. The installed base is smaller than Philips' and MAUDE holds only about a dozen Kodama reports. ACIST's other recalls concern contrast injectors and are excluded. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.