Model note · Otoscope
Riester ri-scope L Otoscope
The ri-scope L is Riester's modular otoscope family. The L1 has direct lighting, while the L2 and L3 use fibre optics and the L3 adds a speculum ejector. Heads fit Riester type C or AA handles, run on 2.5 V alkaline cells or 3.5 V rechargeables with rheotronic dimming, and take XL xenon or LED lamps. A swivel 3x lens moves aside so instruments can be passed into the canal.
The model
Maker, model and type.
- Maker
- Rudolf Riester GmbH
- Model
- ri-scope L Otoscope (L1, L2, L3)
- Device type
- Otoscope
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Variants | L1 direct illumination; L2 and L3 fibre optic; L3 with blue speculum eject button | Source |
| Magnification | Swivel lens about 3x, rotates 360 degrees; separate operation lens for instrumentation | Source |
| Lamp data | XL 2.5 V 750 mA and XL 3.5 V 720 mA, average life 15 h; LED 280 mA, average life 10,000 h | Source |
| Handles and power | Type C or AA rheotronic handles: 2.5 V on alkaline cells, 3.5 V on ri-accu L or ri-accu USB rechargeables | Source |
| Light control | rheotronic dimming by tapping the switch ring; starts at 100% each time; automatic shut off after 180 seconds | Source |
| ri-accu USB indicator | Green steady: full; green flashing: charging; orange: too low, charge now | Source |
| Operating conditions | 0 to 40 C, 30 to 70% RH non condensing, 800 to 1100 hPa | Source |
| Warranty | 2 years from purchase, excluding wear parts | Source |
Recalls
Recalls.
This note lists no recalls. Check the FDA recall database for the latest record.
Safety
Safety warnings from the record.
- Never charge a handle that may have liquid or condensation inside; it can cause a life threatening shock, especially plug in handles. Source
- Do not immerse heads or handles, and do not sterilise or machine reprocess them. Source
- Socket handles must not stay in the socket for more than 24 hours, and the handle must be out of the socket when changing the battery. Source
- When examining while charging a ri-accu USB handle, use only Riester's medical power supply (art. 10709). Source
Sources
Where this note comes from.
- Riester ri-scope L instructions for use (multilingual, copy on tigermedical.com, RIE3779)
- openFDA recall search, Riester
Not covered: The manual copy is a reseller hosted file of unknown revision. The manual's lamp table lists the 3.5 V LED lamp at 2.5 V, which looks like a typing error; only the current and life figures are used. openFDA has no Riester recalls or MAUDE records.