Skip to content

Model note · Hyperthermia System

Pyrexar BSD-2000 Hyperthermia System

The BSD-2000 is a deep regional hyperthermia system that heats deep-seated solid tumours with radiofrequency energy from a ring of antennas surrounding the patient. In the US it is approved under a Humanitarian Device Exemption alongside radiotherapy for cervical cancer patients who cannot have chemotherapy, and it has been sold internationally since 1988. Its key engineering idea is phased-array focusing: frequency, phase and amplitude across several channels are adjusted to steer the heating to the tumour.

The model

Maker, model and type.

Maker
Pyrexar Medical, Inc. (formerly BSD Medical Corporation)
Model
BSD-2000 Hyperthermia System (BSD-2000B and BSD-2000U)
Device type
Hyperthermia System
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
RF frequency range75 to 120 MHzSource
Maximum output power1300 W (BSD-2000B basic) or 1800 W (BSD-2000U upgraded)Source
SubsystemsRF power generator, thermistor-based thermometry, computerised monitoring and control, applicator with patient supportSource
ApplicatorDeep-heating Sigma Base with antenna array applicatorsSource
QA accessoryTissue-equivalent lamp phantom to verify energy focus, steering and operationSource
Site requirementRF-shielded treatment room (not supplied with the system) for FCC complianceSource
Maximum patient body width49 cm left to right (wider is a contraindication)Source
FDA statusHDE H090002, approved 18 November 2011; HUD designation 09-0214Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Contraindicated with pacemakers, defibrillators, pumps, neurostimulators, cochlear implants or any implanted or worn active electronic device. Source
  • Contraindicated with conductive, metal or foreign objects in, on or attached to the body. Source
  • Patients who cannot feel or report pain properly, such as those with significant neuropathy, must not be treated. Source
  • Recognised adverse effects include burns, blistering, ulceration, fat or muscle necrosis and thermal stress. Source
  • Use is restricted to qualified operators supervised by a physician experienced in clinical hyperthermia. Source

Sources

Where this note comes from.

Not covered: The Pyrexar website blocked automated access and no public operator or service manual was found, so error codes, the number of RF channels and antennas, and applicator details are missing. There are no FDA recalls and no MAUDE reports under this brand, so no failing parts could be sourced.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.