Olympus EVIS EXERA III GIF-HQ190 gastrointestinal videoscope

The GIF-HQ190 is a high-definition flexible upper GI videoscope (gastroscope) used in endoscopy units for the oesophagus, stomach and duodenum, running on the EVIS EXERA III CV-190 video centre and CLV-190 light source. Its distinguishing feature is Dual Focus, which switches between normal and near focus for close inspection of mucosa, together with NBI narrow band imaging.

The model

Maker, model and type.

Maker
Olympus Medical Systems Corp.
Model
GIF-HQ190 gastrointestinal videoscope
Device type
Endoscope
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Insertion tube and distal end diameter9.9 mmSource
Working channel2.8 mmSource
Working length103 cmSource
BendingUp 210, down 90, left and right 100 degreesSource
Field of view140 degreesSource
Depth of fieldNormal 5 to 100 mm; near 2 to 6 mm (Dual Focus)Source
FeaturesWater jet, NBI, waterproof one-touch connectorSource
Compatible processorsCV-190 and CV-1500 video system centresSource
Regulatory classFDA Class II, 21 CFR 876.1500; cleared within the EVIS EXERA III Video System (K112680, K131780)Source

Recalls

Known FDA recalls.

  • Z-1275-20222022-04-22

    An A-rubber for a different scope model may have been fitted to GIF-HQ190 during Olympus repair, fitting too tightly.

    FDA record
  • Z-0334-20242023-09-25

    Unvalidated channel air drying after repair; some returned scopes had wet channels, with infection risk.

    FDA record
  • Z-0417-20242023-10-16

    Some repaired GIF-HQ190 scopes lacked adhesive securing the suction channel mouthpiece at the light guide connector.

    FDA record

Safety

Safety warnings from the record.

  • Water left in channels after repair could lead to contamination and infection; a recall found Olympus had not validated its channel drying step after repair. Source
  • Routine reprocessing of a leaking scope can cause further damage, so follow the specific Olympus procedure for leaking scopes. Source
  • If the maintenance unit or light source is switched off while a leaking scope is immersed, water can enter and damage it. Source
  • Olympus returns scopes unrepaired if they have not been cleaned and high-level disinfected or sterilised. Source

Sources

Where this note comes from.

Not covered: The GIF-HQ190 operation and reprocessing manuals were not found on an Olympus public site, so the full reprocessing steps, angulation checks and CV-190 processor error messages are not included.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.