EVIS EXERA III CF-HQ190L colonovideoscope

The CF-HQ190L is a reusable flexible video colonoscope from the Olympus EVIS EXERA III 190 series, used in endoscopy units for examination and treatment of the lower digestive tract from the anus to the ileocaecal valve. It is distinctive for its Dual Focus optics, which switch between a normal and a near focus range at the press of a button, and for built in ScopeGuide coils that show the scope's shape in 3D so loops can be spotted during insertion.

The model

Maker, model and type.

Maker
Olympus Medical Systems Corp.
Model
CF-HQ190L colonovideoscope
Device type
Colonoscope
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Field of view170 degrees in normal focus, 160 degrees in near focusSource
Depth of fieldNormal 5 to 100 mm, near 2 to 6 mm; forward viewingSource
Distal end outer diameter13.2 mmSource
Insertion tube outer diameter12.8 mmSource
Working length1680 mm (L version); total length 2005 mmSource
Instrument channel inner diameter3.7 mm; minimum visible distance 4.0 mm from the distal end in normal focusSource
Angulation rangeUp 180, down 180, right 160, left 160 degreesSource
Compatible systemCV-190 video system centre and CLV-190 xenon light sourceSource
FDA clearance510(k) K131780 (special 510(k), decision 29 August 2013), predicate K112680; class II, product codes FDF, FDS, NWBSource
Sterility and reuseReusable, supplied non-sterile; high-level disinfection or ethylene oxide before useSource

Recalls

Known FDA recalls.

  • Z-0329-20242023-09-25

    Channel air drying after repair was not validated and a small share of repaired scopes went back to customers with wet channels.

    FDA record
  • Z-0415-20242023-10-16

    Improperly repaired light guide connector missing adhesive on the suction channel mouthpiece.

    FDA record
  • Z-1207-20242024-01-11

    Repaired colonovideoscopes assembled without a protective adhesive.

    FDA record

Safety

Safety warnings from the record.

  • Routine reprocessing of a scope that has failed the leak test can drive fluid inside and cause further damage. Use the Olympus leaking scope procedure instead. Source
  • Switching off the maintenance unit or light source while a leaking scope is still under water can let water into internal spaces. Source
  • Water left in channels after repair can support contamination and lead to patient infection. Source
  • The scope is supplied non-sterile and must be high-level disinfected or sterilised before use. Source

Sources

Where this note comes from.

Not covered: The CF-HQ190L operation and reprocessing manuals are not freely downloadable, so model specific inspection steps, leak tester settings and processor messages could not be confirmed; no public error code list exists for the scope itself.