Nihon Kohden Neurofax EEG-1200 series

The Neurofax EEG-1200 is Nihon Kohden's modular digital EEG and sleep study system for routine EEG, long term epilepsy monitoring and ICU work. A system is built from an electrode junction box (head box) at the patient, an isolation unit and main unit that supply isolated power, a Windows PC unit and acquisition and review software. The common JE-921A junction box offers 25 EEG inputs plus extra, bipolar, DC, SpO2 and CO2 inputs, while the JE-120A box raises the count to 256 channels at up to 10 kHz for research and intracranial recording.

The model

Maker, model and type.

Maker
Nihon Kohden Corporation (US sponsor Nihon Kohden America, Inc.)
Model
Neurofax EEG-1200 series (EEG-1200A) with JE-921A or JE-120A electrode junction box
Device type
EEG
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Channel count (JE-921A junction box)25 EEG inputs, 4 extra inputs, 7 bipolar pairs (extra jacks usable as bipolar), 4 DC inputs, 1 SpO2 and 1 CO2 connectorSource
Input impedance and noise (JE-921A)100 Mohm EEG and extra inputs; internal noise 1.5 uVp-p or less (0.53 to 60 Hz)Source
Common mode rejection (JE-921A)105 dB or greater at 60 Hz on EEG inputs; 100 dB or greater on bipolar inputsSource
Digitising (JE-921A)16 bit (97 nV/LSB); all electrodes sampled at once; 100, 200, 500 or 1000 Hz selectable; offset tolerance plus or minus 600 mVSource
Hardware filters (JE-921A)Low cut 0.08 Hz (time constant 2 s); high cut 300 Hz at -18 dB/octSource
High channel junction box (JE-120A)Up to 256 inputs, 24 bit A/D, sampling up to 10 kHz, 200 Mohm input impedance, CMRR 110 dB or more, offset tolerance plus or minus 1000 mVSource
Impedance checkPer electrode impedance on a head layout with thresholds of 2, 5, 10, 20 or 50 kohm; junction box LEDs light for electrodes over thresholdSource
Calibration signal0.25 Hz step or 10 Hz sine wave, 2 to 1000 uV stepsSource
Photic stimulatorMaximum flash energy 1.28 J single; 0.5 Hz, 1 to 33 Hz in 1 Hz steps, 50 and 60 HzSource
Power and safety class750 VA for PC unit and display, 1 kVA with photic and analogue output units; Class ISource
FDA clearanceK080546, EEG-1200A Series Neurofax, Class II (21 CFR 882.1400), decided 29 August 2008; predicate EEG-1100A (K992742); JE-120A box added under K113117Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Outside a medical facility the system may only be used under the direct supervision of a medical professional. Source
  • The system is Class I equipment (protection against shock relies on protective earth), and the datasheet lists it for continuous operation. Source
  • Mains powered parts of the system are fed through the main unit's isolated power supply rather than from separate wall sockets. Source

Sources

Where this note comes from.

Not covered: No original operator or service manual for the EEG-1200 was publicly reachable (copies sit behind paid or upload sites), so there are no error messages, troubleshooting tables, maintenance intervals or failing part data; errorCodes and commonFailingParts are empty for that reason. The datasheet's two column layout extracts in a jumbled order, so only values whose labels were unambiguous were used. No FDA recall names the EEG-1200 or K080546; the only Nihon Kohden EEG product code recall found concerns the PSG-1100 sleep system and was left out. MAUDE has about a dozen Neurofax reports, too few to show a trend. The successor EEG-1260A (K251366) was not covered.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.