Model note · EMG

Natus Viking EMG on the Nicolet EDX platform

Viking is Natus' long running electrodiagnostic line, now sold as the Nicolet EDX system, used in neurophysiology and neuromuscular clinics for nerve conduction studies, needle EMG and evoked potentials, and in theatres for intraoperative monitoring. The platform pairs a base console with built in electrical and auditory stimulators, a 2 or 8 channel isolated amplifier and a PC running the Natus software, which handles tests such as F waves, H reflexes, single fibre EMG and repetitive nerve stimulation.

The model

Maker, model and type.

Maker
Natus Neurology Incorporated, Middleton, Wisconsin (formerly CareFusion / Nicolet)
Model
Natus Viking EMG on the Nicolet EDX platform (earlier Viking Select and Viking Quest systems)
Device type
EMG
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Amplifier channels1 to 8 isolated channels (2 or 8 channel amplifier)Source
CMRRAbove 110 dB, 115 dB typicalSource
NoiseBelow 0.6 uV RMS from 2 Hz to 10 kHzSource
Input impedanceAbove 1000 megohmSource
FiltersLow filter 0.05 Hz to 5 kHz; high filter 30 Hz to 20 kHz; 50/60 Hz notchSource
Conversion24 bit A/D; 384 kHz cumulative samplingSource
Electrical stimulator1 or 2 stimulators, constant current or constant voltage, up to 100 mA or 400 V, 0.01 to 1 ms, single or train, biphasicSource
Auditory stimulatorClick, pip or burst, 0 to 139 dB pSPL, 125 to 8000 HzSource
Impedance meter500 ohm to 480 kilohmSource
Power and isolationMains 100 to 240 V AC, through an isolation transformer depending on configuration; optic or transformer patient isolationSource
FDA clearance510(k) K243982 (Nicolet EDX with Natus Elite), decision 22 January 2025; predicate K120979 (CareFusion Nicolet EDX, 2012); primary product code GWFSource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Labelling limits use in the presence of flammable anaesthetics and with defibrillation equipment. Source
  • Patients have been burned at stimulating or needle electrode sites, including during single fibre EMG and during long procedures with electrosurgery. Source
  • The electrical stimulator output reaches 100 mA in constant current mode or 400 V in constant voltage mode, so output settings carry real burn and discomfort risk. Source
  • The system gives no diagnostic conclusion and is for qualified healthcare providers only. Source

Sources

Where this note comes from.

Not covered: No FDA recall was found for the Natus Viking or Nicolet EDX (Viking recalls in the database are for unrelated patient lifts and catheters), the Viking user guide is only on a third party library behind a terms agreement, and no public error message list was found.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.