Model note · Dermatoscope

HEINE DELTA 30 dermatoscope

The HEINE DELTA 30 is a handheld, battery powered dermatoscope for examining pigmented and other skin lesions in dermatology clinics and primary care. It gives a 10 times magnified image and switches between polarised light, which needs no contact fluid, and non polarised light used with immersion oil. It can be used in contact with an autoclavable scaled contact plate or without contact using an optional working ring, and a smartphone adapter allows image documentation.

The model

Maker, model and type.

Maker
HEINE Optotechnik GmbH & Co. KG
Model
DELTA 30 dermatoscope (K-230.28.305)
Device type
Dermatoscope
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Magnification10xSource
Illumination modesPolarised and non polarised, switched by one button; 3 brightness levelsSource
Power supplyUSB 2.0 Micro B, 5 V DC, 1.2 A, or HEINE Charger 30 station; maximum 6 WSource
BatteryRechargeable lithium ionSource
Charging and operating timeTypically 120 min to charge; typically 160 min of use at highest brightness in polarised modeSource
Electrical classClass II while charging, internally powered in use; contact plate and working ring are type BF applied partsSource
Photobiological safetyRisk group 1 to IEC 62471Source
Operating environment10 to 35 C, 30 to 75% RH, 700 to 1060 hPaSource
Expected service lifeUp to 7 years with designated useSource
Catalogue numberK-230.28.305 (standard edition)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Charge the device at least 1.5 m from the patient environment. Source
  • Examine wounds only in non contact mode; the small contact plate is not for mucosa. Source
  • Do not use in oxygen enriched or anaesthetic environments, or take it into MRI rooms or other strong magnetic fields. Source
  • Do not look straight into the light, and do not use the dermatoscope for eye examination. Source
  • During smartphone image capture, the phone must not have a conductive connection to a PC and mains power; patient data protection on the phone is the user's responsibility. Source

Sources

Where this note comes from.

Not covered: No FDA 510(k), MAUDE reports or recalls were found for the DELTA 30, and HEINE publishes no error codes or service manual, so failing parts are inferred only from IFU warnings. Field of view (often quoted by retailers as 30 mm through a 32 mm lens) was not confirmed on a HEINE page and is left out. Accessory part numbers in the IFU table could not be matched reliably to their items and are left out.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.