Model note · Lithotripter (ESWL)

Dornier Delta III lithotripter

The Dornier Delta III is a modular urology workstation for extracorporeal shock wave lithotripsy (ESWL) of kidney and upper ureteral stones, also used for other imaging guided urological procedures. It combines an electromagnetic shock wave emitter (EMSE) on a therapy arm, an integrated X-ray C-arm and an optional ultrasound arm with a motorised urology table. A camera in the therapy head lets staff see air inclusions in the coupling interface between the treatment head and the patient.

The model

Maker, model and type.

Maker
Dornier MedTech (Dornier MedTech America, Inc.)
Model
Delta III lithotripter
Device type
Lithotripter (ESWL)
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
IndicationFragmentation of renal calyceal, renal pelvic and upper ureteral stonesSource
ModulesThree modules: a basic unit carrying both the X-ray C-arm and the camera equipped therapy arm, a urology patient table, and a control desk with image storage (UIMS)Source
Shock wave sourceElectromagnetic shock wave emitter (EMSE); EMSE 180 on current unitsSource
Penetration depth170 mm (EMSE 180)Source
X-ray generator15 kW, up from 3.5 kW on the Compact Delta predicateSource
ImagingX-ray, ultrasound or both; optional SonoScape X3 Pro ultrasound added in the 2020 clearance; DICOM 3 PACS connectionSource
Therapy head accessOver table and under table positions, more than 120 degrees of accessSource
StandardsIEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-36 (lithotripsy), 60601-2-28, 60601-2-54, IEC 62366-1Source
FDA clearancesK170122 (2017), K172084 (2017), K201074 (2020); Delta III Pro K221903 (2023); product code LNSSource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Haematoma is the dominant reported patient problem (167 MAUDE reports) and Dornier lists it as a potential complication in the operating manual; in the reports sampled, field service usually found the device within specification. Source
  • In one reported haematoma case, service found the F2 focus 2 mm outside Dornier's specified position and realigned it, which shows why the daily alignment check matters. Source
  • The system includes a diagnostic X-ray unit and must be operated under X-ray radiation protection rules (IEC 60601-1-3). Source

Sources

Where this note comes from.

Not covered: The Delta III operating and service manuals are not public, so there are no error codes, energy levels, focal zone dimensions or shock rate limits; acoustic output data was not in the 510(k) summaries reviewed. No FDA recalls were found for the Delta III (the only Dornier recall found concerned laser fibres). The EMSE 180 and 170 mm figures come from Dornier's current page and may apply only to newer units.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.