D.O.R.C. EVA NEXUS Ophthalmic Surgical System

EVA NEXUS is DORC's combined anterior and posterior segment surgical console, the successor to EVA. It uses plunger pumps acting on a cartridge for both infusion and aspiration, so the surgeon can work in flow or vacuum control, and its Smart IOP mode sets infusion pressure from the chosen aspiration and instrument data. With TDC two dimensional cutters that cut on both strokes, it reaches 20,000 cuts per minute. Options include phaco, diathermy, endoillumination and a 532 nm laser.

The model

Maker, model and type.

Maker
D.O.R.C. Dutch Ophthalmic Research Center (International) B.V.
Model
EVA NEXUS Ophthalmic Surgical System (9000.ANTxx / COMyy)
Device type
Vitrectomy System
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Vitrectomy driveUp to 10,000 cycles per minute; 20,000 cuts per minute with TDC cutters (9268 series)Source
AspirationPlunger pump; flow 0 to 90 mL/min or vacuum 0 to -680 mmHg at sea level; rise time 300 msSource
Infusion pressure0 to 150 mmHg; modes Fixed, AIC and Smart IOPSource
Phaco40 kHz, 50 W plus or minus 20 percent, stroke 80 micrometres with 3002.M handpieceSource
Diathermy1 MHz, maximum 10 W at 100 ohmSource
Endoillumination40 lm with 20 ga fibre; spectrum 425 to 680 nm with UV and IR filtersSource
Laser (optional)532 nm frequency doubled Nd:YAG, 0.05 to 1.2 W, 50 to 5000 ms; aiming beam 630 to 645 nm under 1 mW; Class IVSource
TDC cutter sizes23G 0.64 mm, 25G 0.53 mm, 27G 0.43 mm shaftsSource
FDA clearanceK213467, product codes HQC and HQF, June 2022Source

Recalls

Known FDA recalls.

  • Z-2322-20262026-04-30

    TDC VELOCE 23G cutters (9268.VIT23 and packs): loose outer knife complaints; 25G and 27G covered by Z-2323 and Z-2324.

    FDA record

Safety

Safety warnings from the record.

  • The optional laser is Class IV; standard ophthalmic laser safety precautions apply. Source
  • A cutter outer knife that loosens during surgery is a serious intraoperative hazard; quarantine affected lots. Source

Sources

Where this note comes from.

Not covered: DORC's user and service manuals were not publicly reachable, so there are no error codes, maintenance intervals or electrical specifications beyond the 510(k). Other DORC recalls (laser probes with Alcon/Lumenis connectors, a VGPC pack for the earlier EVA) were left out as they do not clearly apply to EVA NEXUS. The manufacturerUrl is DORC's home page and was not opened.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.