Baxter Homechoice Claria APD System

The Homechoice Claria is an automated peritoneal dialysis (APD) cycler that runs overnight exchanges for adult and paediatric patients at home, and can also be used in hospitals. It cycles prescribed volumes of PD solution through a disposable cassette and links to the Sharesource platform, so clinicians can review treatment data and send prescription changes that the patient must accept on the cycler before they take effect.

The model

Maker, model and type.

Maker
Baxter Healthcare Corporation (kidney care business now operating as Vantive)
Model
Homechoice Claria APD System with Sharesource
Device type
Peritoneal Dialysis
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Product codes5C6M40 (Claria APD System, US 510(k)); 5C6M00 (230 V) and 5C6M10 (115/230 V) listed in a Baxter field noticeSource
Sharesource product code5CGM01Source
Patients and settingAdult and paediatric renal failure patients, home healthcare environment and comparable use in professional facilitiesSource
Display and controlsTwo-line OLED screen, control buttons, on-screen step instructions and auto-dim for night useSource
ProgramsFour therapy programs, chosen by the clinician in the Sharesource clinical portalSource
Remote programmingTwo-way, not real time; patient must review and accept changes before the cycler uses themSource
Regulatory classFDA Class II peritoneal dialysis system, 21 CFR 876.5630, product code FKX; predicate HomeChoice and HomeChoice PRO (K102936)Source

Recalls

Known FDA recalls.

  • Z-0409-20262025-08-28

    Sharesource Claria software defect could stop updated prescription settings being saved on the patient's cycler.

    FDA record
  • BfArM 07312/17 (Baxter FCA-2017-017)2017-07-25

    Field notice: cassettes damaged when sets are opened with tools may have undetected holes, risking air delivery to the patient.

    FDA record

Safety

Safety warnings from the record.

  • An undetected cassette hole can deliver air at 10 to 30 mL/min during fill or dwell, which may cause painful pneumoperitoneum. Source
  • Sharesource is not a real-time monitor and cannot programme the cycler in real time; it is not a substitute for clinical management. Source
  • Improper or incorrect procedure is the most common specific problem term in MAUDE reports for this cycler, so user training and technique are worth checking in any incident review. Source

Sources

Where this note comes from.

Not covered: The Claria Patient At-Home Guide, the alarm update clinic letter and any service manual could not be opened (Baxter links now redirect to Vantive), so alarm messages, fill volume ranges, heater temperature, power and battery data are not included.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.