Model note · Glucometer
Abbott Precision Xceed Pro Blood Glucose and B-Ketone Monitoring System
The Precision Xceed Pro is a hospital point of care meter that measures blood glucose and beta hydroxybutyrate (ketone) on individually foil wrapped, bar coded strips. A built in bar code scanner reads operator, patient and strip lot IDs, and results go through a dock to Abbott's PrecisionWeb data management system and on to the LIS. QC lockouts help enforce testing policy.
The model
Maker, model and type.
- Maker
- Abbott Diabetes Care, Inc.
- Model
- Precision Xceed Pro Blood Glucose and B-Ketone Monitoring System
- Device type
- Glucometer
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Measurement principle | Amperometric biosensor; glucose by GDH-NAD with 1,10-phenanthroline quinone mediator; ketone by hydroxybutyrate dehydrogenase | Source |
| Test time | Glucose 20 seconds; ketone 10 seconds | Source |
| Measuring range | Glucose 20 to 500 mg/dL; ketone 0.0 to 8.0 mmol/L (Hi shown above 8.0) | Source |
| Haematocrit limits | Glucose 20 to 70%; ketone 30 to 60% | Source |
| Sample volume and types | Glucose 0.6 uL, ketone 1.5 uL; capillary, venous, arterial and neonatal whole blood (venous and arterial within 30 minutes of collection) | Source |
| Power | Two AA alkaline or NiMH rechargeable cells | Source |
| Size and weight | 19.7 x 7.5 x 5.33 cm; 256 g | Source |
| Memory | 2,500 patient results, 1,000 control results, 6,000 operators, 6,000 patient IDs, 18 glucose strip lots | Source |
| Strip storage | Shelf life 15 to 18 months at 4 to 30 C | Source |
| FDA clearance | K080960 (ketone capability added, September 2008); meter and glucose strips from K021960; product codes NBW, LFR, JIN | Source |
Recalls
Known FDA recalls.
- Z-0994-20112010-12-22
Certain Precision Xceed Pro (and related Abbott) glucose strip lots could give erroneously low results.
FDA record
Safety
Safety warnings from the record.
- Not for patients who are dehydrated, in shock, critically ill or hyperosmolar, and not for screening or diagnosing diabetes. Source
- Falsely low results from affected strip lots could lead to wrong treatment decisions. Source
- Neonatal samples are allowed, so confirm haematocrit sits within the 20 to 70% glucose limit before relying on a result. Source
Sources
Where this note comes from.
- FDA 510(k) decision summary K080960, Precision Xceed Pro BGKMS
- CAP Today product guide, Bedside glucose testing systems, March 2013 (vendor supplied data)
- openFDA recall search, Xceed Pro
- openFDA MAUDE problem counts, Precision Xceed Pro
Not covered: No public copy of the operator's manual could be opened (a training site link returned 404), so error codes, QC rules, cleaning agents and environmental limits are missing. Many specs come from a 2013 CAP Today guide filled in by Abbott, not an Abbott document. Current market status was not confirmed.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.